Promotion Restriction
Not eligible for promotion
Temp. Range
35 Degrees to 375 Degrees C, 95 Degrees to 707 Degrees F
Stem Hole Dia.
0.0625 in, 0.125 in, 0.15625 in, 0.1875 in and 0.25 in Probes
Temp. Stability
+/-0.07 Degrees C at 50 Degrees C; +/-0.1 Degrees C at 100 Degrees C; +/-0.3 Degrees C at 375 Degrees C
Temp. Accuracy
+/- 0.25 Degrees C at 100 Degrees C, +/- 0.25 Degrees C at 50 Degrees C, +/- 0.5 Degrees C at 375 Degrees C, +/-0.45 Degrees F at 122 Degrees F, +/-0.9 Degrees F at 707 Degrees F
Calibration Certificate
NIST
Power Required
115V AC ( +/-10%), 55/65 Hz
With Process Connections
No
Stability
+/-0.07 Degrees C at 50 Degrees C, +/-0.1 Degrees C at 100 Degrees C, +/-0.3 Degrees C at 375 Degrees C
Display Resolution
0.1 Degrees
Hot Block Temperature Range
95 to 707 F, 35 to 375 C
Overall Dimensions
2 1/4 in H x 4 29/32 in W x 6 7/64 in D
Typical Uniformity
+/-0.2 Degrees C
Temperature Accuracy
+/-0.25 Degrees C (at 100 Degrees C), +/-0.25 Degrees C (at 50 Degrees C), +/-0.45 Degrees F (at 122 Degrees F), +/-0.5 Degrees C (at 375 Degrees C), +/-0.9 Degrees F (at 707 Degrees F)
Calibrator Type
Handheld Dry-Well
Calibration Details
Data at 50 Degrees C, 100 Degrees C, 150 Degrees C, 200 Degrees C, 250 Degrees C, 300 Degrees C and 375 Degrees C
Typical Cooling Time
14 min (375 Degrees to 100 Degrees C)
Brand and Series
Fluke 9100S
For Stem Diameter
0.0625 in, 0.125 in, 0.15625 in, 0.1875 in, 0.25 in
Calibrator Accuracy
+/-0.25 Degrees C at 100 Degrees C, +/-0.25 Degrees C at 50 Degrees C, +/-0.5 Degrees C at 375 Degrees C
Includes NIST-Traceable Certificate
Yes
Set Point Style
Adjustable
Typical Heating Time
9.5 min (35 Degrees to 375 Degrees C)
Power Source Details
115V AC, 230V AC
Temperature Range
95 Degrees to 707 Degrees F/35 Degrees to 375 Degrees C
Country of Origin (subject to change)
United States