| Property | Value |
|---|---|
| Promotion Restriction | Not eligible for promotion |
| Item | Drywell |
| Loop Supply Voltage | 24V DC |
| Thermocouple Type | C, E, J, K, L, M, N, R, S, T, U |
| Accuracy | +/-0.03 Degrees C |
| RTD Type | NI-120, 385, 3892, JIS |
| RTD Temperature Range | -196 to 660 Degrees C |
| Thermocouple Temperature Range | 0 to 1480 Degrees C (type S) |
| Display | LCD |
| Includes | 9143 Field Metrology Well, Insert for Configuration Ordered, Power Cord, RS-232 Cable, User Guide, Technical Manual CD, Report of Calibration, Calibration Label, 6-pin DIN Connector, Test Lead Kit, Clamp-on ferrites (3), Tongs (insert removal tool), 9930 Interface-it Software and User's Guide |
| Interface | RS-232 |
| Calibration Certificate | NVLAP accredited |
| Agency Compliance | CE, CSA, C-TICK, ROHS, WEEE |
| Max. Loop Current | 24 mA |
| Standards | NVLAP Accredited |
| Thermocouple mV Characteristics | 0.025% + 0.01 mV at -10 to 75 mV |
| Application | Field Testing |
| For Stem Diameter | (2) 3 mm, (2) 4 mm, (2) 6 mm |
| Brand and Series | Fluke 9143 |
| Display Temperature Accuracy | +/-0.2 Degrees C |
| For Thermocouple Type | Type C, Type E, Type J, Type K, Type L, Type M, Type N, Type R, Type S, Type T, Type U |
| Well Depth | 5.91 in |
| Well Size | 5 29/32 D Well |
| Calibrator Type | Field Metrology Well |
| Reference Junction Temperature Characteristics | +/- 0.35 Degrees C Accuracy at Ambient Temperature of 13 Degrees to 33 Degrees C |
| Overall Height | 11.41 in |
| Overall Depth | 11.61 in |
| Calibrator Weight | 16 lb |
| RTD Temperature Range C | -196 Degrees to 660 Degrees C |
| Overall Width | 7.31 in |
| Current mA Characteristics | 0.02% + 2 mV Accuracy/4 to 22 mA, 4 to 24 mA Range |
| Reference Input Temperature Characteristics | +/-0.010 Degrees C at -25 Degrees C, +/-0.015 Degrees C at 0 Degrees C, +/-0.020 Degrees C at 50 Degrees C, +/-0.025 Degrees C at 150 Degrees C, +/-0.030 Degrees C at 200 Degrees C, +/-0.040 Degrees C at 350 Degrees C, +/-0.050 Degrees C at 420 Degrees C, +/-0.070 Degrees C at 660 Degrees C |
| Stability | +/-0.02 Degrees C at 200 Degrees C, +/-0.02 Degrees C at 33 Degrees C, +/-0.03 Degrees C at 350 Degrees C |
| With Process Connections | Yes |
| Thermocouple Temperature Accuracy Characteristics | Type C: +/-1.1 Degrees C at 660 Degrees C, Type E: +/-0.7 Degrees C at 660 Degrees C, Type J: +/-0.7 Degrees C at 660 Degrees C, Type K: +/-0.8 Degrees C at 660 Degrees C, Type L: +/-0.7 Degrees C at 660 Degrees C, Type M: +/-1.4 Degrees C at 660 Degrees C, Type N: +/-0.9 Degrees C at 660 Degrees C, Type R: +/-1.4 Degrees C at 660 Degrees C, Type S: +/-1.5 Degrees C at 660 Degrees C, Type T: +/-0.8 Degrees C at 400 Degrees C, Type U: +/-0.75 Degrees C at 600 Degrees C |
| Power Source Details | 100 to 115V, 230V |
| For RTD Type | 4-Wire, Ni-120 |
| Typical Uniformity | +/-0.04 Degrees C at 33 Degrees C, +/-0.1 Degrees C at 200 Degrees C, +/-0.2 Degrees C at 350 Degrees C, +/-0.01 Degrees C at 33 Degrees C, +/-0.015 Degrees C at 200 Degrees C, +/-0.02 Degrees C at 350 Degrees C |
| Calibrator Accuracy | +/-0.2 Degrees C (Display - Full Range) |
| Includes NIST-Traceable Certificate | No |
| Thermocouple Temperature Range C | 0 Degrees to 1,480 Degrees C |
| Resistance Sensing Current Characteristics | 0.02% at 2 mV; Cal 4-22 mA, Spec 4-24 mA |
| Display Resolution | 0.01 Degrees |
| Calibration Details | NVLAP Accredited Calibration Included |
| Reference Input Resistance Characteristics | +/--0.0025 ohm Accuracy (0 to 42 ohm), +/-60 ppm of Reading Accuracy (42 to 400 ohm) |
| Typical Cooling Time | 32 min (350 Degrees to 33 Degrees C), 14 min (350 Degrees to 100 Degrees C) |
| Temperature Range | -196 Degrees to 660 Degrees C, 0 Degrees to 1,480 Degrees C |
| Probe Connection Types | (1) 4-Terminal RTD Input, (1) Small Connectors TC |
| Typical Heating Time | 5 min (33 Degrees to 350 Degrees C) |
| Overall Dimensions | 11 13/32 in H x 7 5/16 in W x 11 39/64 in D |
| Calibrator Application | Industrial Process Environment |
| Sensor Type | RTD, Thermocouple |
| Set Point Style | Adjustable |
| Number of Set Points | Variable |
| 4-Wire Device Resistance Characteristics | Range: 0 to 400 ohms/Accuracy: +/-0.0025 ohms (0 to 42 ohms), +/-60 ppm of Reading (42 to 400 ohms) |
| Single Well/Dual Well | Single Well |
| Temperature Accuracy | +/-0.03 Degrees C |
| Country of Origin (subject to change) | United States |
| Package Quantity | 1 |